Flow cytometry / Flow antibody |
大小 | |
Width×depth (不含MACS MiniSampler) | 60.5×40 cm |
Width×depth (含MACS MiniSampler) | 60.5×50 cm |
Height (可調整觸控螢幕高度) | 39.3–52 cm |
Laser | Channel | Filter (nm) | Dye or parameter |
紫光雷射405 nm (40mW diode) |
V1 V2 |
450/50 525/50 |
CFP, VioBlue VioFreen |
藍光雷射488 nm (30mW DPSS) |
B1 B2 B3 B4 |
525/50 585/40 655-730 750 LP |
GFP, FITC PE PI, PerCP, PE-Cy5.5 PE-Vio770, PE-Cy7 |
紅光雷射635 nm (20 mW diode) |
R1 R2 |
655-730 750LP |
APC APC-Vio770, APC-Cy7 |
藍光雷射488 nm | FSC SSC |
488/10 488/10 |
Size Granularity |
Laser | Channel | Filter (nm) | Dye or parameter |
紫光雷射405 nm (40mW diode) |
V1 V2 |
450/50 525/50 |
CFP, VioBlue VioGreen |
藍光雷射488 nm (50 mW DPPSS) |
B1 B2 |
525/50 614/50 |
GFP, FITC, YFP PI, Lss-mKate |
黃光雷射561 nm (100 mW diode) |
Y1 Y2 Y3 Y4 |
586/15 615/20 661/20 750LP |
PE, dsRed mCherry, TexasRed PE-Cy5, APC, mKate PE-cy7, APC-cy7, PE-Vio770, APC-Vio770 |
5. 自動細胞標記 (Auto-labeling) MACSQuant Analyzer 10與MACSQuant VYB具有自動細胞標記染色的功能,是目前在市場上的流式細胞儀之中唯一具備的能力。透過條碼讀取器來輸入所使用的試劑資料並選取樣本管 (最多可以進行96個樣本標記),軟體有能力自動進行樣本染色程序,在樣本與樣本染色之間有清洗探針的功能,讓您的樣本不會交叉汙染。有效的減少手動時間、減少人為誤差和增加數據的再現性與可信度。 |
6. 自動化程式 MACSQuant是全自動化的流式細胞儀,具有各種自動化的功能,包含: 啟動與暖機程序、校正、樣本標記、清洗探針、螢光補償、流體管路清洗與關機程序。 |
最小吸取量 | 25 uL (建議) |
最小樣本大小 | 1 um (建議) |
樣本流速 | 流速25-100 uL/minute,加上自動化維持流速在每秒500、1000、2000個事件 |
最大事件流速 | 最高可到每秒104事件 |
最高的測量速度 | 30分鐘分析一個96孔盤中的樣本 (96個樣本,每個樣本體積10 uL) |
絕對細胞計數 | 精準度+/-5%;再現性 (CV)<5% |
樣本吸取 | 樣本吸取口呈現y/z軸的運動方式,並有自動清洗功能 |
樣本殘留量 | 標準值0.01% |
![]() |
![]() |
MACSQuant Analyzer的絕對細胞計數是出色的 且再現性高。每毫升所含的Counting Beads數量, 經由廠商提供的數據表(A),由MACSQuant Analyzer 計算五管Counting Beads的平均數值(B)。結果顯 MACSQuant Analyzer 計算出的結果與廠商數據 沒有任何差異。 |
將細胞做序列稀釋(1, 0.5, 0.25, 0.1, 0.05倍),經由 MACSQuant測定可得到每毫升的絕對細胞數, 並且數值是成線性關係的。 |
Pre-enrichment的優點 | WorkFlow |
|
![]() |
MACSQuant流式細胞分析儀10
![]() |
|
MACSQuant流式細胞分析儀VYB
![]() |
|
Accessories 配件
![]() |
Bioprocess impurities |
Mycoplasma detection |